WRIGHTS LANE SYNTHES USA PRODUCTS LLC NORIAN DRILLABLE INJECT 5CC-STERILE; FILLER, CALCIUM SULFATE PREFORMED PELLETS Back to Search Results
Model Number 07.704.005S
Medical Device Problem Code Failure to Eject (4010)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial reporter is synthes sales representative.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.A review of the device history records has been requested.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2018, the norian drillable inject was found disconnected from the package.Specifically, the delivery syringe was detached from the product pouch.The product was wasted and never made it into the sterile field while mixing at the back table.It is unknown if there was a surgical delay.Patient and procedure outcome are unknown.This report is for one (1) norian drillable inject 5cc-sterile.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device is not expected to be returned for manufacturer review/investigation.Device history records review was completed for part: 07.704.005s, lot: dse6832.Manufacturing location: monument, release to warehouse date: jan 30, 2018, expiry date: aug 28, 2019.Purchased finished goods traveler met all inspection acceptance criteria.Inspection sheet, incoming final inspection met all inspection acceptance criteria.Certificate of conformance received from dsm dated 22-dec-2017 was reviewed and determined to be conforming.Certificate of conformance sterility requirements and results were reviewed and determined to be conforming.This lot met all dimensional, visual, sterility and packaging criteria at the time of release with no issues documented during the inspection or release of the product that would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NORIAN DRILLABLE INJECT 5CC-STERILE
Common Device Name
FILLER, CALCIUM SULFATE PREFORMED PELLETS
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key8111620
Report Number2939274-2018-55127
Device Sequence Number2261884
Product Code MQV
UDI-Device Identifier10886982131614
UDI-Public(01)10886982131614
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K102722
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/28/2019
Device Model Number07.704.005S
Device Catalogue Number07.704.005S
Device Lot NumberDSE6632
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/09/2018
Supplement Date Received by Manufacturer01/01/2019
Initial Report FDA Received Date11/28/2018
Supplement Report FDA Received Date01/02/2019
Patient Sequence Number1
-
-