| Catalog Number |
03L80-21 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The customer reported of falsely depressed architect carbon dioxide results for multiple samples.The customer indicated the initial results were approximately 18 mmol/l and retested on a new lot with a results of approximately 22 mmol/l.There was no reported impact to patient management.
|
| |
|
Additional Manufacturer Narrative
|
|
Review of complaint activity determined that there were no other complaints related to falsely depressed results for the likely cause lot.Tracking and trending report review for the carbon dioxide assay determined that there were no related adverse or non-statistical trends.The customer's instrument logs were reviewed.This review found the absorbances for the last calibration with the likely cause lots were lower than previous calibrations, which may indicate possible contamination of the carbon dioxide reagent.Using world wide field data the performance of reagent lot 53082uq05 and calibrator lot 11449fd01 were reviewed and no issues were identified.Manufacturing documentation for the assay was reviewed and did not identify any issues.Labeling was reviewed and sufficiently addresses the customer's issue.No systemic issue or deficiency of the architect carbon dioxide assay was identified.
|
| |
|