(b)(4).Date sent to the fda: 01/02/2019.Investigation summary: it was reported packaging labeling identification.An opened box of product code z452, lot # mgm577 with fourteen unopened samples of product code z332, lot # jgk489 and an unopened sample of product code z452, lot mgm577 were returned for analysis.This box requires twelve samples per box.During the visual inspection of sample, a mismatched between box (z452, lot # mgm577) and fourteen foils (code z332, lot # jgk489) were observed.However, the manufactured dates were reviewed and there is a difference of three years of separation between the manufacture date of the lot # jgk489 (date of mfg 6/23/2015) and lot # mgm577 (date of mfg 6/6/2018).The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this batch.According to manufactured dates there are a difference in the years of the manufacture of the lots of product and that the box was received opened, the reported complaint of " suture inside of box is labeled z332h and outside of box says z452g." it could not be happened in our manufacturing sites and the reported complaint was due to external cause.The device history records were reviewed for lot mgm577 and the manufacturing criteria was met prior to the release of this lot.Product code z452g, expiration date: 05/31/2023, date of mfg.: 06/06/2018.The device history records were reviewed for lot jgk489 and the manufacturing criteria was met prior to the release of this lot.Product code z332h, expiration date: 05/31/2020, date of mfg.: 06/23/2015.
|