STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE 132 SIZE 2.5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
|
Back to Search Results |
|
| Catalog Number |
6020-2530 |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Codes |
Inflammation (1932); Tissue Damage (2104); Injury (2348); Reaction (2414); Fluid Discharge (2686)
|
| Date of Event |
03/19/2012
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Product not available.
|
| |
|
Event or Problem Description
|
|
Its reported by the husband, that his wife underwent a total hip arthroplasty on (b)(6) 2012.It is further alleged that the hip failed and she is scheduled for revision on (b)(6) 2018.Update (b)(6) 2018 - as reported by rep via fax, patient was revised (b)(6) 2018: "patient was brought in for revision surgery on right hip.[surgeon] debrided copious amounts of soft tissue, damaged from what he described as metal debris.[surgeon] removed the liner and head and noted black metal debris present on.Trunnion and inside head.Patient had a lot of discolored fluid and tissue removed.[surgeon] deemed the abductors insufficient from the damage and was forced to implant a constrained liner for stability.".
|
| |
|