Model Number MICL13.2
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Codes Intraocular Pressure Increased (1937); No Code Available (3191)
Date of Event 10/25/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
The reporter indicated the surgeon implanted a 13.2mm micl13.2 implantable collamer lens, -10.0 diopter, in the patients right eye (od) on (b)(6) 2018.The lens was explanted on (b)(6) 2018 due to excessive vaulting, which caused an increase in pressure.The lens was exchanged for a shorter lens and the problem was resolved.
 
Additional Manufacturer Narrative
Corrected data: previous mdr stated: "excessive vault which caused an increase in pressure".Corrected to : "excessive vault only" in the initial mdr.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key8113339
Report Number2023826-2018-01752
Device Sequence Number10824195
Product Code MTA
UDI-Device Identifier00841542103299
UDI-Public00841542103299
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2019
Device Model NumberMICL13.2
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/01/2018
Supplement Date Received by Manufacturer01/02/2019
Initial Report FDA Received Date11/28/2018
Supplement Report FDA Received Date01/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL AND LOT NUMBER UNK; INJECTOR MODEL AND LOT NUMBER UNK
Outcome Attributed to Adverse Event Required Intervention;
Patient Age30 YR
-
-