| Model Number |
MMT-7020A |
| Medical Device Problem Codes |
Incorrect, Inadequate or Imprecise Result or Readings (1535); Calibration Problem (2890); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
Rash (2033)
|
| Date of Event |
11/18/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the sensor had inaccurate readings that triggered threshold suspend alarm.The customer¿s blood glucose was 101 mg/dl and the sensor glucose was 92 mg/dl.Insulin delivery was suspended due to sensor values.The customer was informed that their blood glucose and sensor glucose levels were not in acceptable rang.The sensor will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Inspected one opened and sensor and performed continuity resistance test.Sensor passed per specifications.Sensor initialization test was performed using a new lab transmitter with a paradigm pump.Connected sensor to the transmitter and placed it in 100 bbs solution and confirmed communication icon was displayed and that the transmitter was flashing while connected to the sensor.The sensor functioned properly.Cannot perform buffered test solution, the sensor is beyond its expiration date.Unable to confirm adhesive anomaly on opened/used sensor.(b)(4).
|
| |
|