| Catalog Number |
G7783515 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Neurological Deficit/Dysfunction (1982)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the r eported event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was pre-operatively diagnosed with cervical spondylotic radiculopathy and underwent anterior fusion at c4/5/6.Post-op, screw breakage at c6 was reported.Dysraphism and recurrence of symptoms were reported as the result of the event.Posterior cervical decompression and fusion was performed additionally on (b)(6) 2018.
|
| |
|
Additional Manufacturer Narrative
|
|
Radiographic image review: post-op x-ray of ct for c4-6 acdf show broken screw at c6.There is peek interbody graft at both ends, bone growth through the c4-5 graft is present but not so at c5-6.There is bone growth into the graft but not entirely through it.If information is provided in the future, a supplemental report will be issued.
|
| |
|