JOHNSON AND JOHNSON SURGICAL VISION, INC. TECNIS CL; MONOFOCAL IOLS Back to Search Results
Model Number Z9002
Medical Device Problem Codes Failure to Eject (4010); Separation Problem (4043)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 10/30/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Age/date of birth: unknown/ not provided.If implanted, give date: not applicable, as lens was removed/replaced during the same procedure.If explanted, give date: not applicable, as lens was removed/replaced during the same procedure.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the trailing haptic of a model z9002, 20.0 diopter intraocular lens (iol), when being inserted into the eye, came off.The trailing haptic got caught and stayed in the cartridge as the lens was placed in the eye.There was an incision enlargement and sutures.Doctor used the back up lens and the surgery went well with no patient injury reported.
 
Additional Manufacturer Narrative
The packaging box was received.The package inside had a cartridge with an empty lens case and a detached haptic.No lens was received.The cartridge received was observed with scarce number of viscoelastic residues just in the cartridge tube.Since a haptic was received detached the complaint haptic detached was verified, although is unknown if the haptic was from the lens because it was not received.The stuck in cartridge complaint was not verified.The manufacturing purchase orders were evaluated, and the devices were manufactured within specifications.No non- conformance reports were found.The units were released according to specification.Complaint history review was performed and it revealed that no other complaint has been received for this production order number.The labeling review was completed and revealed that the directions for use (dfu) provide the customer with proper usage instructions and guidelines.No labeling changes are required.As a result of the investigation there is no indication of a product quality deficiency.An attempt has been made to obtain missing information; however, no definitive response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS CL
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON AND JOHNSON SURGICAL VISION, INC.
1700 east st. andrew place
santa ana CA 92705
MDR Report Key8113630
Report Number2648035-2018-01527
Device Sequence Number16709587
Product Code HQL
UDI-Device Identifier05050474530690
UDI-Public(01)05050474530690(17)220307
Combination Product (Y/N)Y
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P880081
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/05/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/07/2022
Device Model NumberZ9002
Device Catalogue NumberZ900200200
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/29/2018
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 10/31/2018
Supplement Date Received by Manufacturer12/20/2018
10/25/2020
Initial Report FDA Received Date11/28/2018
Supplement Report FDA Received Date12/28/2018
11/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL PSCST30, LOT UNKNOWN
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/01/2005
-
-