| Model Number |
PCB00 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Codes |
Visual Disturbances (2140); No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown, not provided, but the best estimate date is between (b)(6) 2018.(b)(4).Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that tecnis preloaded monofocal intraocular lens (model pcb00, +16.5 diopter) was explanted from patient¿s right eye in a secondary procedure because of patient experiencing specific visual issues.Reportedly lens with same model, +25.0 diopter, a larger diopter was used as replacement for the patient.Patient was doing good.No further information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation: yes, returned to manufacturer on: 11/29/2018, device returned to manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.The pcb00 piece model was returned in a plastic container.Visual inspection using microscope magnification was performed: the returned pcb00 lens was received cut in pieces.The condition in which the sample returned is consistent with a product that was explanted.Due the condition of the pcb00 model received the complaint issue reported as visual disturbance was not possible to be confirmed.The unit were released according to specification.A product quality deficiency was not identified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search revealed that no other complaints were received for this production order.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa:(b)(4).
|
| |
|