Model Number 24667
Medical Device Problem Code Sensing Intermittently (1558)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The returned device was evaluated.External visual inspection showed no damage.When connected to a root both channels of the module were able to take measurements which were consistent with measurements from a known good module.Humidity was added to the module and the measured values were no longer consistent with a known good module.Internal inspection showed the cable connector had contamination damage.The sensor connector was temporarily replaced with a known good connector and the module was fully functional.A service history record review reveals that this unit was in the field for over one (1) year with no previous reported issues related to this reported event.
 
Event or Problem Description
The customer reported the module was plugged into moc-9 and not displaying anything on root monitor.Also on previous occasion stopped displaying during surgery.No patient impact or consequences were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O3 REGIONAL OXIMETER MODULE
Common Device Name
OXIMETER
Manufacturer (Section D)
MASIMO - 40 PARKER
40 parker
irvine CA 92618 1604
MDR Report Key8113792
Report Number2031172-2018-00560
Device Sequence Number13883579
Product Code MUD
UDI-Device Identifier00843997010771
UDI-Public00843997010771
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K160526
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 11/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number24667
Device Catalogue Number9637
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/20/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/09/2018
Initial Report FDA Received Date11/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-