| Model Number |
24667 |
| Medical Device Problem Code |
Sensing Intermittently (1558)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The returned device was evaluated.External visual inspection showed no damage.When connected to a root both channels of the module were able to take measurements which were consistent with measurements from a known good module.Humidity was added to the module and the measured values were no longer consistent with a known good module.Internal inspection showed the cable connector had contamination damage.The sensor connector was temporarily replaced with a known good connector and the module was fully functional.A service history record review reveals that this unit was in the field for over one (1) year with no previous reported issues related to this reported event.
|
| |
|
Event or Problem Description
|
|
The customer reported the module was plugged into moc-9 and not displaying anything on root monitor.Also on previous occasion stopped displaying during surgery.No patient impact or consequences were reported.
|
| |
|