| Model Number |
88901 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/14/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Attempts to obtain patient information have been unsuccessful.No tests/laboratory data was available.The implant or explant dates are not applicable to this device.
|
| |
|
Event or Problem Description
|
|
This case was reviewed and investigated according to the manufacturer's policy.It was reported that during a diagnostic procedure the image was lost.The procedure was successfully completed with a second of the same make and model.Visual, microscopic inspection, and functional testing were performed on the returned device.The pim connector was separated and not returned with the device.Additionally, the tip was deformed and the device was missing small portions of adhesive.The probable cause of the reported failure could not be determined due to the pim connector being separated and not returned with the device.The probable cause of the observed missing adhesive could not be conclusively determined by the investigation.The manufacturing documentation for this device was reviewed and the device met all quality and manufacturing release criteria.This product problem is being submitted because adhesive was missing from the device, it could not conclusively be determined when the separation occurred.There is a potential for harm if the malfunction were to recur.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 other text : placeholder.
|
| |
|
Event or Problem Description
|
|
This case was reviewed and investigated according to the manufacturer''s policy.Internal reference: (b)(4).This follow-up supplemental report #1 is being submitted to report additional investigation findings, to wit: there is no potential for harm if the malfunction were to recur.This complaint was reassessed based on an engineering study that indicated, this family of catheters do not exhibit any adhesive detachment during normal use.There is no evidence of abnormal use conditions, it is likely the adhesive detached post-procedure, during the process of returning the device for analysis.There was no patient injury, no status decline, no adverse event, and no unplanned additional medical or surgical intervention reported.There is no potential for harm if the malfunction were to recur.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to the alleged death or deterioration in the state of the health of any person.
|
| |
|
Additional Manufacturer Narrative
|
|
Supplemental report #1 indicated that the adhesive likely detached post procedure; however, the conclusion code was not updated from 4315 (cause not established) to 4308 (cause traced to transport/storage).
|
| |
|