MAUDE Adverse Event Report: US SURGICAL PUERTO RICO VERSASTEP; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

US SURGICAL PUERTO RICO VERSASTEP; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number VS101500
Medical Device Problem Codes Break (1069); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during a robotic hysterectomy procedure, the surgeon misaligned the trocar and pierced the sleeve on the side.A piece of the cannula broke off and fell into the patient's cavity but was retrieved by a grasper.They replaced the sleeve to re-enter the abdomen.No patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection noted: the distal end of the expandable sleeve.The end is frayed, also there is a trocar inserted through the sleeve, the end of the trocar is gouged.The obturator appeared intact.A functional evaluation found that the device passed an air leak test.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the gouged cannula tip may occur if the device is mishandled or handled roughly during application.The root cause of the observed damage was misuse of the product which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSASTEP
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
MDR Report Key8113926
Report Number2647580-2018-05813
Device Sequence Number1770994
Product Code GCJ
UDI-Device Identifier10884521084155
UDI-Public10884521084155
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K012539
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2021
Device Model NumberVS101500
Device Catalogue NumberVS101500
Device Lot NumberP6H0043X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/26/2018
Initial Date Received by Manufacturer 11/05/2018
Supplement Date Received by Manufacturer02/07/2019
Initial Report FDA Received Date11/28/2018
Supplement Report FDA Received Date03/04/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age35 YR
-
-