Model Number 8886848882 |
Device Problems
Device Difficult to Setup or Prepare (1487); Failure to Fire (2610)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/05/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during a laparoscopic proctectomy procedure, the two devices were unable to clip.The external body did not hit after the internal track was applied.A new product was used to complete the case.There was no patient injury.
|
|
Manufacturer Narrative
|
Evaluation summary: post market vigilance (pmv) led an evaluation of three devices.The visual inspection noted the three cartridges were received applied with the clip bodies disengaged and the tracks were not inserted properly at the distal end of the clip bodies.Functionally; the clip bodies and tracks were reloaded into the three cartridges and fired with acceptable results.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the observed condition may occur if the clip was not properly applied and mishandled during the clinical application.The root cause of the observed damage was misuse of the product which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
|
|