At this time this issue is still being investigated in (b)(4) and a post market product issue assessment, (b)(4), is currently being conducted.When more information becomes available, a supplemental will be submitted.
|
|
Merge cardio is a system intended to be used to acquire, store, print, transfer, and archive clinical information including images, hemodynamic studies and reports, measurements (via import from dicom structured reporting, text files or optical character recognition of measurements captured on images) and cardiology signal (waveform) data.On (b)(4) 2018, it was found internally while patching the customers 3.5 reporting knowledge base to 5.0 reporting knowledge base that some of the michigan z-scores were calculating incorrectly.This is due to the formula using 'b' vs 'bsa' for body surface area.This variable has been causing z-scores to calculate incorrectly.Due to an incorrect value displaying in the diagnostic report, there is a potential for incorrect treatment of a patient that could result in harm.However at this time, there have been no allegations of patient harm as a result of this issue.(b)(4).
|