| Model Number |
170043 |
| Medical Device Problem Codes |
Break (1069); Component or Accessory Incompatibility (2897); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/06/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic nissen fundoplication procedure, while loading the device, the needle broke off from the instrument tip.There was no patient injury.
|
| |
|
Search Alerts/Recalls
|