• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO MEDICAL CORPORATION RADIFOCUS INTRODUCER II KIT; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO MEDICAL CORPORATION RADIFOCUS INTRODUCER II KIT; INTRODUCER, CATHETER Back to Search Results
Model Number N/A
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/17/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient identifier - requested, not provided.Age & date of birth - requested, not provided.Sex - requested, not provided.Weight - requested, not provided.Ethnicity - requested, not provided.Race - requested, not provided.Implanted date: device was not implanted.Explanted date: device was not explanted.The actual device has not been returned for evaluation.The investigation is currently ongoing.A follow up report will be submitted once the investigation is complete.For this reason, evaluation code 11 has been referenced in the conclusions.A review of the device history record of the product code/lot# combination was conducted with no findings.
 
Event or Problem Description
The user facility reported that upon insertion of the ik sheath, it rolled up on itself.
 
Additional Manufacturer Narrative
This report is being submitted as follow up no.1 provide additional information and to update.Based on the additional information received and the evaluation results, the reported event has been deemed not an fda reportable event.
 
Event or Problem Description
Additional information was received on december 5, 2018.The procedure being performed was a right hearth catheterization.The dilator was fine, however, the end of the sheath rolled up like an accordion upon insertion.The sheath was removed and a new one of the same kind was used.The patient was stable trough the whole procedure and the procedure outcome was successful.A 5f mak micro-puncture was used 1st to dilator skin and blood vessel.There was no surgical or medical intervention required and there was no blood loss.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RADIFOCUS INTRODUCER II KIT
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
TERUMO MEDICAL CORPORATION
950 elkton blvd.
elkton MD 21921
MDR Report Key8116022
Report Number1118880-2018-00176
Device Sequence Number1770431
Product Code DYB
UDI-Device Identifier00389701007496
UDI-Public00389701007496
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K954234
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2020
Device Model NumberN/A
Device Catalogue NumberRSS802
Device Lot NumberWF16
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/29/2018
Initial Date Received by Manufacturer 11/14/2018
Supplement Date Received by Manufacturer11/29/2018
Initial Report FDA Received Date11/29/2018
Supplement Report FDA Received Date12/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-