| Model Number |
87011 |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/30/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No evaluation possible.The instruments have not been returned for evaluation nor has the identifying lot number(s) been provided.Upon the receipt of additional information and/or the product in question, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
While inserting a screw into the iliac, the tips of both arsenal screwdrivers snapped and broke.The case was successfully completed.No patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation of the returned driver found that the distal tip of the instrument had fractured and became separated.Inspection under magnification revealed circular fracture lines indicative of a torsional overload.
|
| |
|