| Catalog Number |
M653G |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
09/28/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.This medwatch report is in response to receipt of maude event report #mw5080384.Additional information was requested and the following was obtained: what measures were taken to retrieve the broken piece? the surgeon attempted to remove the broken needle tip with his instruments but was unable to do so.Was there any additional tissue damage as a result of searching for the needle piece? no additional tissue damage noted.Were x-rays taken to locate the needle piece(s)? x-ray confirmed the needle tip in in the sternum.At what location was the needle found? the patient¿s sternum.Was the patient brought back to or to have needle removed or was the needle removed during the initial procedure? no.If retained, what is the surgeon's opinion of consequences to the patient? the patient cannot have an mri in the future.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a coronary artery bypass graft procedure on (b)(6) 2018, and suture was used.During closure of the patient¿s chest the needle broke off in the patient¿s sternum.Attempts to retrieve the fragment were unsuccessful.No patient consequence was reported.
|
| |
|