(b)(4).Concomitant medical products: biomet microfixation tmj system right standard mandibular component 50 mm, catalog #: 24-6550, lot #: 530070a.Biomet microfixation tmj system left fossa component, medium, catalog #: 24-6561, lot #: 547160b.Biomet microfixation tmj system right fossa component, medium, catalog #: 24-6560, lot #: 516330b.Biomet microfixation 2.4 mm system high torque (ht) cross-drive screw 2.7 x 8 mm, catalog #: 91-2708, lot #: ni.Biomet microfixation 2.4 mm system high torque (ht) cross-drive screw 2.7 x 10 mm, catalog #: 91-2710, lot #: ni.Biomet microfixation tmj system cross drive fossa screw 2.0 x 7 mm, catalog #: 99-6577, lot #: ni.Biomet microfixation tmj system cross drive emergency fossa screw 2.3 x 9 mm, catalog #: 99-6589, lot #: ni.Biomet microfixation 2.7 mm system emergency cross drive screw 3.2 x 10 mm, catalog #: 99-9950, lot #: ni.Therapy date: unknown.Customer has indicated that the product will not be returned to zimmer biomet for investigation; it remains implanted in the patient.Reported event was unable to be confirmed due to limited information received from the customer.The product identity was confirmed through tmj tracking.No product was returned and no functional tests or inspections could be performed.No x-rays, scans, pictures, or physician's reports were provided.For aforementioned reasons, and due to the fact that pain is a relative condition, the complaint cannot be verified and the root cause was unable to be determined.The instructions for use for this product warns of the potential for complications.It states in the section titled adverse events: facial swelling and/or pain, facial nerve dysfunction.It also states in the section titled patient counseling information: discussion of the following points is recommended prior to surgery.The risks associated with a total tmj system (see warnings and adverse events).Post-operative pain relief and return of function varies from patient to patient.Device history record (dhr) was reviewed and no discrepancies were found.There are no indications of manufacturing defects.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.Multiple mdr reports were filed for this event; please see associated reports: 0001032347-2018-00811 through 0001032347-2018-00819.
|
|
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing of the mandible and fossa components.Lot identification is necessary for review of device history records, lot identification was not provided for the screws.A definitive root cause cannot be determined for the left side implants.No problem was found with the right side implants.Pain was reported to be experienced on the left side, so this issue is related to the left side implants and no allegations were made against the right side implants.The reported event is unconfirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.This report is being submitted to update additional information in sections b4, b5, d4, g3, g6, h2, h3, h6 and h11.
|