|
It was reported that the stent inadvertently deployed when attempting to remove it from being stuck to the wire.A 6x60x130 innova vascular stent was selected for use for a stenting procedure in the superficial femoral artery (sfa).During preparation, outside of the patient's body, the physician attempted to load the stent over an unspecified.035 guidewire but the stent became stuck on the guidewire.When the physician attempted to pull the stent off of the wire, the stent started to inadvertently deploy.It partially deployed.Another innova vascular stent was used to complete the procedure.There were no patient complications reported.
|
|
It was reported that the stent inadvertently deployed when attempting to remove it from being stuck to the wire.A 6x60x130 innova vascular stent was selected for use for a stenting procedure in the superficial femoral artery (sfa).During preparation, outside of the patient's body, the physician attempted to load the stent over an unspecified.035 guidewire but the stent became stuck on the guidewire.When the physician attempted to pull the stent off of the wire, the stent started to inadvertently deploy.It partially deployed.Another innova vascular stent was used to complete the procedure.There were no patient complications reported.Device evaluated by mfr: the outer sheath, tip, inner sheath and the remainder of the device were checked for damage.Visual examination revealed a kink at the nosecone.Kinks at the nosecone most likely occurs from handling, when it is packaged to be returned for investigation.The yellow thumbwheel protector is still attached but is on backwards.The pull grip is pulled 33mm out and there is blood present in the inner liner.The stent was completely outside of the device.Microscopic examination revealed that there is foreign material (fm) present on the stent.An amplatz super stiff.035 guidewire was used and the guidewire could pass through the entire device.Inspection of the remainder of the device presented no other damage or irregularities.
|