Catalog Number TF0183511
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 11/09/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records were provided to the manufacturer.An image was provided.The lot number for the device was provided.The device history records and image are currently under review.The device was not returned to the manufacturer for evaluation.The investigation is currently under way.
 
Event or Problem Description
It was reported that during an aortic valvuloplasty procedure, the balloon valvuloplasty catheter allegedly got caught on the guidewire during advancement.It was further reported that after significant resistance the balloon valvuloplasty catheter was exchanged over the guidewire for another and the procedure was successfully completed.There was no patient contact.
 
Event or Problem Description
It was reported that during an aortic valvuloplasty procedure, the balloon valvuloplasty catheter allegedly got caught on the guidewire during advancement.It was further reported that after significant resistance the balloon valvuloplasty catheter was exchanged over the guidewire for another and the procedure was successfully completed.There was no patient contact.
 
Additional Manufacturer Narrative
The device history records have been reviewed with special attention to the raw materials, subassemblies, manufacturing process, and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported to date for this lot number and failure mode.The investigation is inconclusive, as the sample was not returned for evaluation.The root cause could not be determined based upon available information.It is unknown whether procedural issues contributed to the event.The review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUE FLOW DILATATION CATHETER
Common Device Name
BALLOON VALVULOPLASTY CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key8118514
Report Number2020394-2018-02141
Device Sequence Number1710676
Product Code OZT
UDI-Device Identifier00801741097645
UDI-Public(01)00801741097645
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K172932
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/28/2017
Device Catalogue NumberTF0183511
Device Lot NumberGFAS2140
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/11/2016
Supplement Date Received by Manufacturer12/20/2018
Initial Report FDA Received Date11/29/2018
Supplement Report FDA Received Date01/16/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-