Model Number DR80510
Medical Device Problem Codes Break (1069); Leak/Splash (1354)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/27/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.
 
Event or Problem Description
It was reported that during an angioplasty procedure an alleged leak was found in the catheter shaft of the pta balloon during the first inflation attempt.There was no reported patient injury.
 
Event or Problem Description
It was reported that during an angioplasty procedure an alleged leak was found in the catheter shaft of the pta balloon during the first inflation attempt.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the device was returned for evaluation.The device was inflated with water, and a leak was identified at the proximal balloon butt joint.Upon closer examination, a partial break at the butt joint was identified, resulting in the leak.Therefore, the investigation is confirmed for the reported leak, as well as for a break at the butt joint.The partial break at the glue butt joint resulted in the identified leak.However, the definitive root cause for the partial break could not be determined based upon available information.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORADO PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key8118524
Report Number2020394-2018-02137
Device Sequence Number19253120
Product Code DQY
UDI-Device Identifier00801741058578
UDI-Public(01)00801741058578
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K072283
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDR80510
Device Catalogue NumberDR80510
Device Lot Number93YB0069
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2018
Initial Date Received by Manufacturer 11/05/2018
Supplement Date Received by Manufacturer03/26/2019
Initial Report FDA Received Date11/29/2018
Supplement Report FDA Received Date04/02/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age70 YR
-
-