• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. RESERVOIR 1.8ML MMT-326A; PUMP, INFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. RESERVOIR 1.8ML MMT-326A; PUMP, INFUSION Back to Search Results
Model Number MMT-326A
Medical Device Problem Code Improper Flow or Infusion (2954)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event or Problem Description
The customer reported via phone call the insulin pump alarmed motor error and no delivery.The customer¿s blood glucose level was 109 mg/dl at the time of the incident.The customer stated that the insulin pump.The customer stated that the drive support cap was normal and that the insulin pump was not exposed to high magnetic fields and was able to complete the rewind process.Customer also reported air bubbles in the reservoir.Customer reported that they did not allow for insulin to warm to room temperature prior to filling the reservoir.The customer was advised to discontinue use of the insulin pump and to revert to the backup plan.The insulin pump is being replaced and is expected to return for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESERVOIR 1.8ML MMT-326A
Common Device Name
PUMP, INFUSION
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key8118672
Report Number2032227-2018-68961
Device Sequence Number10829183
Product Code FRN
UDI-Device Identifier00643169771413
UDI-Public(01)00643169771413(17)210512(10)HG2JF9T
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/12/2021
Device Model NumberMMT-326A
Device Catalogue NumberMMT-326A
Device Lot NumberHG2JF9T
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/21/2018
Initial Report FDA Received Date11/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age68 YR
Patient Weight160
-
-