Model Number ZKB00
Medical Device Problem Codes Break (1069); Separation Problem (4043)
Health Effect - Clinical Codes Corneal Edema (1791); No Code Available (3191)
Date of Event 11/06/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Implant and explant dates: if implanted or explanted; give date: not applicable as the lens was inserted and removed.(b)(4).All pertinent information available to johnson and johnson surgical vision has been submitted.
 
Event or Problem Description
It was reported that the haptic broke off during insertion of tecnis multifocal intraocular lens (model zkb00, +20.5 diopter).The lens was removed and replaced in the same procedure.Reportedly an incision enlargement was performed during the procedure, but no sutures were used and no vitrectomy was performed.The patient had small amount of edema but doing ok.No further information was provided.
 
Additional Manufacturer Narrative
Device evaluation: the product testing could not be performed as the product was not returned.The reported complaint cannot be confirmed.Manufacturing record review: the manufacturing process record was evaluated and no deviation was found during process related to the complaint issue reported.There was no discrepancy found during the review.The product was manufactured and released according to specifications.A search revealed that no other complaints were received for this production order.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1 MULTIFOCAL
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key8119032
Report Number3011852734-2018-00224
Device Sequence Number18882592
Product Code MFK
UDI-Device Identifier05050474560093
UDI-Public(01)05050474560093(17)220313
Combination Product (Y/N)Y
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/05/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/13/2022
Device Model NumberZKB00
Device Catalogue NumberZKB00U0205
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 11/06/2018
Supplement Date Received by Manufacturer12/17/2018
10/25/2020
Initial Report FDA Received Date11/30/2018
Supplement Report FDA Received Date12/21/2018
11/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
Date Report Sent to FDA01/01/2005
-
-