| Model Number |
799-AEV1-01 |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to treat a patient (age & gender unknown), the device displayed an "power on self test failure - 1084" error message.Complainant indicated that the clinician bag mask ventilated to continue treating the patient.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.Please reference medwatch report 1220908-2018-03482 for a similar event reported from the first device used on the same patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation and the customer's report was not replicated or confirmed.The device was put through extensive testing including bench handling, extensive stress testing, functional testing and a full internal inspection without duplicating the report.The autocal valve was replaced as a precaution.The device was recertified and returned to the customer.No trend is associated with reports of this type.
|
| |
|