MAUDE Adverse Event Report: PRIDE MOBILITY PRODUCTS PRIDE MOBILITY PRODUCTS; POWERED WHEELCHAIR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRIDE MOBILITY PRODUCTS PRIDE MOBILITY PRODUCTS; POWERED WHEELCHAIR Back to Search Results
Model Number JAZZY 614HD
Medical Device Problem Codes Fire (1245); Smoking (1585)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The "date of event" was not provided.The device has not yet been made available for evaluation.If the device or further information becomes available, a follow-up report will then be issued.
 
Event or Problem Description
Alleges chair was charging when he smelled smoke and he noticed it coming from the controller on his chair.Alleges he unplugged the charger and took the controller off of the chair to take it outside.Alleges as he was taking it outside, it caught on fire.Alleges throwing it in his yard and pouring water on it.
 
Additional Manufacturer Narrative
A returned product evaluation was conducted and the thermal event was not related to the product.
 
Event or Problem Description
Alleges chair was charging when he smelled smoke and he noticed it coming from the controller on his chair.Alleges he unplugged the charger and took the controller off of the chair to take it outside.Alleges as he was taking it outside, it caught on fire.Alleges throwing it in his yard and pouring water on it.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIDE MOBILITY PRODUCTS
Common Device Name
POWERED WHEELCHAIR
Manufacturer (Section D)
PRIDE MOBILITY PRODUCTS
401 york ave
duryea PA 18642
MDR Report Key8119642
Report Number2530130-2018-00169
Device Sequence Number13505786
Product Code ITI
UDI-Device Identifier00606509100052
UDI-Public00606509100052
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K042612
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Operator of Device Lay User/Patient
Device Model NumberJAZZY 614HD
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/29/2018
Supplement Date Received by Manufacturer11/29/2018
Initial Report FDA Received Date11/30/2018
Supplement Report FDA Received Date01/10/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-