Model Number TRUETRACK
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/08/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.(b)(4).Manufacturer contacted customer (several attempts) in a follow-up call to ensure that the replacement products resolved the initial concern - unable to establish contact with the customer at this time.
 
Event or Problem Description
Consumer reported complaint for high blood glucose test results.Mother is calling on behalf of the customer.The customer is concerned with tests results from results obtained of 276, 275, 200, 320 and 343 mg/dl.The customer's expected fasting blood glucose test result range is 120 - 165 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.The customer was not available to perform a back to back blood test during the call on (b)(6) 2018.Product is stored according to specification in the bedroom.The test strip lot manufacturer's expiration date is 09/13/2020 and open vial date is 4 - 5 days.The meter memory was reviewed for previous test result history: result 1:276mg/dl date: (b)(6) 2018 time:3:35pm fasting; result 2:275mg/dl date: (b)(6) 2018 time:10:28am fasting; result 3:200mg/dl date: (b)(6) 2018 time:3:45pm fasting; result 4:320mg/dl date:(b)(6) 2018 time:9:25am fasting; result 5:343mg/dl date:(b)(6) 2018 time:8:03pm fasting.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUETRACK
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key8120140
Report Number1000113657-2018-01214
Device Sequence Number229043
Product Code NBW
UDI-Device Identifier00052569132463
UDI-Public(01)00052569132463
Combination Product (Y/N)N
Initial Reporter StateFL
PMA/510(K) Number
K032657
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 11/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/13/2020
Device Model NumberTRUETRACK
Device Catalogue NumberA4H01-81
Device Lot NumberRV5256
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/08/2018
Initial Report FDA Received Date11/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SECOND THERAPY
-
-