• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMERICAN MEDICAL SYSTEMS, INC. SUREFLEX FIBERS; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMERICAN MEDICAL SYSTEMS, INC. SUREFLEX FIBERS; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number S-LLF273TG
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Event date: the exact date of this event is not known.The total length of the fiber is 11 feet 7 inches, connector not included in over-all length.The connector is detached and returned.The fiber proximal end exhibits severe burns and melting.The distal tip of the fiber exhibits no signs of usage.Black discoloration was noted near proximal end within blue jacket.The fiber was tested with hene laser fixture, aim beam is visible at break location.Cannot determine when and how fiber was detached from product after final inspection.
 
Event or Problem Description
It was reported that during a surgical procedure the "lma" was found to be detached.The fiber was not used.It was not reported how the procedure was completed.There was no patient outcome information reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUREFLEX FIBERS
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
AMERICAN MEDICAL SYSTEMS, INC.
3070 orchard drive
san jose CA 95134
MDR Report Key8120318
Report Number2937094-2018-60138
Device Sequence Number1656937
Product Code GEX
UDI-Device Identifier00878953004839
UDI-Public00878953004839
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K140679
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 11/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/06/2021
Device Model NumberS-LLF273TG
Device Catalogue NumberS-LLF273TG
Device Lot Number0021813636
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2018
Initial Report FDA Received Date11/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-