• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THE SPECTRANETICS CORPORATION SPECTRANETICS TIGHTRAIL ROTATING DILATOR SHEATH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THE SPECTRANETICS CORPORATION SPECTRANETICS TIGHTRAIL ROTATING DILATOR SHEATH Back to Search Results
Model Number 545-513
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pulmonary Embolism (1498); Cardiomyopathy (1764); Death (1802); Endocarditis (1834); Sepsis (2067); Perforation of Vessels (2135); Great Vessel Perforation (2152); Iatrogenic Source (2498)
Date of Event 11/02/2018
Type of Reportable Event Death
Additional Manufacturer Narrative
Patient weight was not available from the site.The lot# of the device was not available from the site as they discarded all devices after the case and the philips representative was not at the case.The expiration date is also unavailable as it is determined by the lot#.Device manufactured date cannot be determined as the lot# is unavailable.The device was disposed of after the case.There is no allegation that the spectranetics/philips device caused/contributed to the event.
 
Event or Problem Description
On (b)(6) 2018, a cardiac lead management procedure for the removal of 2 leads (1 right atrial and 1 right ventricular) due to bacteremia was performed.A spectranetics 13fr tightrail rotating dilator sheath 545-513 was being used when a small tear less than 1cm in the patient's superior portion of their superior vena cava (svc) was noted.Rescue efforts were implemented and emergent sternotomy was performed.The patient survived the procedure.However, the patient¿s family removed the patient from life support on (b)(6) 2018.
 
Additional Manufacturer Narrative
Additional information received via phone conversation with manufacturer's representative regarding this event; this information is populated.Additional patient code added.
 
Event or Problem Description
Additional information: post market surveillance analyst spoke by phone with manufacturer's representative on (b)(4) 2018, who was not present at this procedure, to inquire if there were additional circumstances surrounding patient's death, since the reported svc (superior vena cava) injury was reportedly repaired.Representative stated that he had heard, subsequent to complaint submission, that the patient had an azygos injury instead, and emailed the physician for further information as to the details of the event.This additional information was not available and reportedly will not be provided to manufacturer.
 
Additional Manufacturer Narrative
Additional case detail provided by user facility.Relevant history/medical conditions provided by user facility.Spectranetics lead locking device added as concomitant medical product patient codes added.
 
Event or Problem Description
On december 27, 2018, the manufacturer received additional information from the facility via email, providing additional details of this event.There was no alleged reported malfunction of any spectranetics devices in use during the event.The facility director of cardiology reported the patient sustained only an azygos tear, and not an svc tear as initially reported.The two leads planned for extraction were a right atrial (ra) and right ventricular (rv) lead.Both leads reportedly contained spectranetics lead locking devices being used for traction platforms.Only the rv lead was extracted during the case; the ra lead was not extracted; it was abandoned due to case complexity.The suspected cause of the injury was reportedly a combination of the spectranetics tightrail device and pulling forces (spectranetics lld) that were both in use at the time of the event.The model and lot numbers of the lld device present in the rv lead are not available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRANETICS TIGHTRAIL ROTATING DILATOR SHEATH
Common Device Name
TIGHTRAIL
Manufacturer (Section D)
THE SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
MDR Report Key8120618
Report Number1721279-2018-00164
Device Sequence Number197559
Product Code DRE
UDI-Device Identifier00813132020897
UDI-Public00813132020897
Combination Product (Y/N)Y
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K150360
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/05/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number545-513
Device Catalogue Number545-513
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 11/05/2018
Supplement Date Received by Manufacturer12/13/2018
12/27/2018
Initial Report FDA Received Date11/30/2018
Supplement Report FDA Received Date12/19/2018
01/22/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death; Hospitalization; Life Threatening; Required Intervention;
Patient Age68 YR
Date Report Sent to FDA01/01/2005
-
-