• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® X-FORCE® BALLOON DILATION CATHETER; URETERAL BALLOON CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® X-FORCE® BALLOON DILATION CATHETER; URETERAL BALLOON CATHETER Back to Search Results
Catalog Number 997610
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event or Problem Description
It was reported that the surgeon complained that the balloon was not deflating completely when it was time to remove the balloon from the ureter following dilation.Upon testing, the balloon deflated normally.The bard sales rep.Stated the issue stemmed from the fact that the user (the scrub tech) was not waiting long enough for the vacuum effect to take full effect.In order to deflate the balloon, the inflator should be deployed back fully to allow the pressure to pull all of the fluid out of the balloon and deflate it.Final conclusion was that the product was working properly and this event was a user error.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: ¿description the x-force® balloon dilation catheter is a dual lumen catheter with a balloon mounted on the distal tip.It has a radiopaque tip and two radiopaque markers beneath the balloon that define the working length.The lumen labeled with the rated burst pressure (xx atm) is for balloon inflation.The other lumen allows the catheter to track over a 0.038¿ (.97mm) diameter guidewire and can be used for monitoring of pressure or the infusion of medication and/or contrast medium.Each balloon inflates to a stated diameter and length at a specific pressure, typically 10 atm.The x-force® balloon dilation catheter comes packaged with a refolding tool and is available with or without an inflation device.It comes sterile and is for single use only.Indications for use the x-force® balloon dilation catheter is recommended for use in the dilation of the urinary tract.Contraindications do not use the x-force® balloon dilation catheter in the presence of conditions, which create unacceptable risk during the dilation of the urinary tract.Warnings ¿ do not exceed the recommended rated burst pressure (rbp) for this device.Balloon rupture may occur if the rbp rating is exceeded.Please refer to the device label for the recommended maximum rated burst pressure.Extreme caution should be used when considering dilation of irregular, not-compliant tissue or in the presence of certain calculi.¿ this is a single use device.Do not re-sterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Precautions caution: federal law (u.S.A.) restricts this device to sale by or on the order of a physician.Only a physician who has an understanding of the clinical applications, technical principles and associated risks of balloon dilation within the urinary tract should use this device.After use, the x-force® balloon dilation catheter and/or accessories may be considered a potential biohazard.Handle and dispose of in accordance with medical practice and applicable laws and regulations.Inspection prior to use the x-force® balloon dilation catheter is a sterile, single use device.Carefully inspect the catheter and the sterile packaging for signs of damage that may have occurred during shipment.Do not use the product if damage is evident.Preparation of the catheter all x-force® balloon dilation catheters contain air in the balloon lumen.The air must be removed to allow liquid to fill the balloon when it is inflated.1.Remove the protective sheath from the balloon.2.Attach the inflation device to the connector on the balloon lumen.3.Open the stopcock and draw back on the inflation device to remove the air from the balloon catheter.4.Close the stopcock, remove the inflation device, depress the plunger to remove any air and reattach to the balloon catheter.5.Repeat steps 3-4 until all air is removed from the balloon lumen.Catheter insertion 1.Introduce the catheter carefully over a 0.038¿ (.97mm) guidewire and place it in the area that needs to be dilated.Use the radiopaque markers to aid in proper positioning.Note: dilation procedures should be conducted under fluoroscopic guidance with appropriate x-ray equipment or direct vision.Caution: do not advance or withdraw the catheter or guidewire against any significant resistance.The cause of the resistance must be determined fluoroscopically and remedial action taken.X-force® balloon dilation catheter".
 
Event or Problem Description
It was reported that the surgeon complained that the balloon was not deflating completely when it was time to remove the balloon from the ureter following dilation.Upon testing, the balloon deflated normally.The bard sales rep.Stated the issue stemmed from the fact that the user (the scrub tech) was not waiting long enough for the vacuum effect to take full effect.In order to deflate the balloon, the inflator should be deployed back fully to allow the pressure to pull all of the fluid out of the balloon and deflate it.Final conclusion was that the product was working properly and this event was a user error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® X-FORCE® BALLOON DILATION CATHETER
Common Device Name
URETERAL BALLOON CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key8120912
Report Number1018233-2018-05697
Device Sequence Number252652
Product Code EZN
UDI-Device Identifier00801741052118
UDI-Public(01)00801741052118
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K050875
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number997610
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/06/2018
Supplement Date Received by Manufacturer01/18/2019
Initial Report FDA Received Date11/30/2018
Supplement Report FDA Received Date02/07/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-