| Model Number |
15 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
Death (1802)
|
| Date of Event |
10/23/2018
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Physio-control evaluated the customer's device and was unable to duplicate the reported issue.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.The cause of the reported issue could not be determine.Physio-control performed a clinical review of the event and concluded that the device use could have contributed to the patient's outcome.Physio-control reviewed the electronic patient record from the event and determined that defibrillation was needed, but the provision of defibrillation was delayed by more than 30 seconds.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that they were using this device on a patient and had an issue with the aed function of the device.The device returned a "no shock advised" message to the user two to three times before advising the user to shock the patient.The user delivered a shock to the patient and then the device went back to the "no shock advised" message.The customer is unsure if the device returned the correct shock advisory messages during the event.The patient involved in this event, an (b)(6) male, is deceased.No further patient information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
A physio-control research scientist evaluated the electronic patient record from the event and determined that the "no shock advised" decisions made by the device during the first three rhythm analyses were due to the presence of steep negative slopes in the ecg rhythm.The results were not indicative of a deviation from the established sensitivity and specificity of the lifepak shock advisory algorithm.No device malfunction was observed during the analysis of the electronic patient record.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that they were using this device on a patient and had an issue with the aed function of the device.The device returned a "no shock advised" message to the user two to three times before advising the user to shock the patient.The user delivered a shock to the patient and then the device went back to the "no shock advised" message.The customer is unsure if the device returned the correct shock advisory messages during the event.The patient involved in this event, an (b)(6) male, is deceased.No further patient information was provided.
|
| |
|