BARD ACCESS SYSTEMS DOT POWERPICC,4SL,3CG; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
|
Back to Search Results |
|
| Model Number |
CK000244 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/08/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a break in the 3cg stylet was confirmed.One 3cg stylet was returned for investigation.A break in the polyimide tubing was observed 1.8 cm from the distal tip of the stylet.A microscopic examination of the adjoining ends of the stylet revealed a rough and uneven surface.The stylet wire exhibited evidence of use.The stylet wire had been partially retracted through the extension set, kinked at the proximal end of the septum, and taped to the extension set.The entire stylet was returned for investigation.Kinks were observed in the polyimide tubing both proximal and distal to the break site.A break in the core wire coincided with the break in the polyimide tubing.A magnet was missing from the distal end of the proximal polyimide tubing segment.The wall thickness of the polyimide tubing was within specification.It was reported that the stylet was bent when it was removed from the catheter.The wire tip broke off in the bag once it was set aside for investigation.A lot history review (lhr) of 18fbe856 showed no other similar product complaint(s) from this lot number.[(b)(4)].
|
| |
|
Event or Problem Description
|
|
It was reported that there was a fractured guidewire.It was stated the wire was bent when it was taken out of the patient.The wire tip broke off in the bag once it was set aside for investigation.Patient anatomy was not difficult.Additional information rec'd via medwatch stated, "single lumen picc trimmed to 37 cm.Introduced to about 15 cm, then noted resistance and picc removed.When inspecting the stylet, it was fractured at 2 cm from the tip.".
|
| |
|
Search Alerts/Recalls
|
|
|