STRYKER TRAUMA KIEL GUIDE WIRE, BALL-TIPPED, STERILE T2 TIBIA Ø3X800 MM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
|
Back to Search Results |
|
| Catalog Number |
18060080S |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/29/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The returned device matched the report and the complaint failure mode was confirmed.Referring to product inquiry the guide wire, ball-tipped, sterile t2 tibia ø3x800 mm [packaging] is stated to be the primary product.No further associated products were reported.A review of the dhr revealed no discrepancies.Sealing seams and original packaging are still intact.Thus, the pollution must have occurred during the packaging process at our sub-supplier, which was not detected during inspection.Based on the above facts the root cause of the reported event is related to an inadequate packaging process / insufficient inspection specification.The found foreign material [like a hair] has to be classified as a non-conformance.Ncr was triggered for further root cause investigation.
|
| |
|
Event or Problem Description
|
|
In in-coming inspection at distribution, it was found that there was a foreign material (hair / bug) in blister.This event was found on 1 of 50pcs of 18060080s, lot# k0b324f received on our po#(b)(4) and 1 of 100pcs of 18060085s, lot# k09a2f3 received on our po#(b)(4).No patient involvement, adverse consequences, delay or medical intervention.
|
| |
|