| Catalog Number |
72202959 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that was not possible to see through device, unable to get focus.There was not delay in case and back-up device was available.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation was performed by the supplier and could confirm the customer complaint for it was not possible to see through the device, unable to get focus.A visual inspection was performed and showed the scope to have distal tip damage, a cracked weld, and cracked internal lenses.This damage is caused by contact with another source.
|
| |
|