MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR Back to Search Results
Model Number 37612
Device Problems Charging Problem (2892); Communication or Transmission Problem (2896); Inadequate Instructions for Non-Healthcare Professional (2956)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/05/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a consumer regarding a patient who was implanted with a neurostimulator (ins) for parkinson's dual and movement disorders.It was reported that the patient said she had a battery charge from a pc to rc and that was due to normal battery depletion.Patient said they were telling her how to recharge when she had the iv in going into surgery and the rest when she was coming out of surgery and she doesn't remember how to charge.Troubleshooting done on the call included re-positioning the antenna over the ins, turning the dial through all 4 positions, and reviewing the antenna locate.Patient could get only 4 bars.They tried the antenna locate and couldn't get the antenna locate to come up.Troubleshooting did not resolve the issue.The patient had bandages present.Once the bandages come off she should be able to get more bars.No symptoms were reported.No further complications were reported or anticipated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key8121952
MDR Text Key129067274
Report Number3004209178-2018-26642
Device Sequence Number1
Product Code MHY
UDI-Device Identifier00643169864238
UDI-Public00643169864238
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 11/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/01/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/28/2019
Device Model Number37612
Device Catalogue Number37612
Was Device Available for Evaluation? No
Date Manufacturer Received11/05/2018
Date Device Manufactured09/13/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age79 YR
-
-