| Model Number |
VFC061020-V |
| Medical Device Problem Codes |
Separation Failure (2547); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/08/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The lot number was provided.A review of the approved device history records indicated the lot met all release criteria.A lot history trending review was performed and there were no similar complaints for this lot number.The device has not been returned to the manufacturer.The instructions for use (ifu) identifies premature of difficult coil detachment as potential complications associated with use of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that during treatment of a cerebral aneurysm, the coil did not detach.During withdrawal, the coil detached in the microcatheter.Both segments of the coil were removed in their entirety together with the microcatheter.There was no reported patient injury or intervention.The patient was reported to be in good health.
|
| |
|
Additional Manufacturer Narrative
|
|
Initial report indicated that the device was returning for manufacturer analysis; however, the device has not been returned for investigation.A product analysis could not be performed.The root cause is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for analysis.The distal loop of the implant was slightly stretched and lost its shape.Several kinks and loss of shape were observed at the middle of the implant.The distal portion of the implant was undamaged.The proximal tie knot was in good condition.The monofilament was not visible in the bottom of the marker band.The implant was opened at the proximal tie knot and the monofilament was noted to be broken at the tie knot glue location.The monofilament break location is indicative of a detachment caused by a tensile break in the monofilament.A potential cause of the damage is over specification tensile forces being placed on the device.The report of non-detachment could not be confirmed, since the pusher was not returned for analysis.
|
| |
|