| Model Number |
961-580 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
11/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No parts have been returned to the manufacturer for analysis.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation device being used outside of a procedure.It was reported that the tracker¿s fixation failed despite turning the screw of the clamp on the tracker.There was no patient present when this issue was observed.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the suretrak tracker clamp was returned for analysis.Through visual/physical examination, the reported issue was able to be duplicated.The adjustment screw of the returned clamp has been cross threaded, damaging the threads of the screw.Analysis determined that there was a mechanical failure with the returned suretrak tracker clamp.This issue was documented in a medtronic navigation hardware anomaly tracking database.If information is provided in the future, a supplemental report will be issued.
|
| |
|