• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION TECNIS SYMFONY TORIC; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION TECNIS SYMFONY TORIC; MULTIFOCAL IOLS Back to Search Results
Model Number ZXT375
Medical Device Problem Code Inaccurate Information (4051)
Health Effect - Clinical Codes Blurred Vision (2137); No Code Available (3191)
Date of Event 10/25/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported zxt375 18.0 diopter intraocular lens (iol) was implanted in the patient¿s ocular sinister (left eye) on (b)(6) 2018 and explanted on (b)(6) 2018 due to refractive surprise.The surgery center contact stated the toric multifocal lens was unnecessary, and induced more cylinder.The zxt375 iol was exchanged for a zlb00 16.5 diopter iol.Through follow up, customer account provided additional information confirming ocular sinister (left eye) is blurry compared to ocular dexter (right eye), blurry issue was first observed (b)(6) 2018, more cylinder power was induced (b)(6) 2018, and clarified the toric multifocal was unnecessary due to inaccurate keratometry readings, toric correction was not indicated once repeat measurements were done.No additional information was provided to johnson & johnson surgical vision.
 
Additional Manufacturer Narrative
Additional information: device available for evaluation ¿ yes, returned to manufacturer on 1/7/2019.Device returned to manufacturer ¿ yes.Device evaluation: the intraocular lens (iol) was returned to the manufacturing site for investigation.The product was received in a little jar with liquid.The product was disinfected.Visual inspection with the unaided eye shows the product is cut in pieces, most probably to make explant possible.Considering the condition of the returned lens, additional analysis is not possible.Manufacturing records review: a review of the manufacturing records was performed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints have been received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION
1700 e. st. andrew place
santa ana CA 92705
MDR Report Key8122767
Report Number9614546-2018-01125
Device Sequence Number295088
Product Code POE
UDI-Device Identifier05050474606937
UDI-Public(01)05050474606937(17)230421
Combination Product (Y/N)Y
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/05/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/21/2023
Device Model NumberZXT375
Device Catalogue NumberZXT375U180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2019
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 11/05/2018
Supplement Date Received by Manufacturer01/14/2019
10/25/2020
Initial Report FDA Received Date12/01/2018
Supplement Report FDA Received Date01/29/2019
11/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age81 YR
Date Report Sent to FDA01/01/2005
-
-