Model Number MMT-522NAH
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/27/2015
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The insulin pump was received with a detached end cap.Unable to perform the displacement test, rewind, basic occlusion test, occlusion test, prime and compromised forced sensor system verify excessive no delivery and motor error alarms due to detached end cap.
 
Event or Problem Description
The customer reported via phone call that their device experience more than one motor error alarm within the last thirty days.Report does not state whether the reservoir is able to lock into place.The customer¿s blood glucose level at the time of the incident is unknown.Customer was advice disconnect from device and to refer to their healthcare provider¿s instructions.No further details or information available.The device will be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-522NAH
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key8123286
Report Number2032227-2018-69463
Device Sequence Number11166812
Product Code OYC
UDI-Device Identifier00613994622686
UDI-Public(01)00613994622686
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-522NAH
Device Catalogue NumberMMT-522NAH
Device Lot NumberA2522NAHJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/23/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/26/2018
Initial Report FDA Received Date12/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age65 YR
Patient Weight60
-
-