| Model Number |
MMT-723RNAP |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/22/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Unable to prime during prime/compromised force sensor system test due to faulty fsr (gold).Pump passed basic occlusion and displacement test.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump had motor error.The customer¿s blood glucose level was unknown at the time of the incident.The drive support cap was normal.Customer was able to successfully clear the alarm.Customer was able to complete insulin pump rewind.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump will be returned for analysis.
|
| |
|