• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. COCR 12/14 FEM HD 22 + 0; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. COCR 12/14 FEM HD 22 + 0; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 71302200
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 09/13/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a revision surgery was performed due to patient (b)(6) suffered a dislocation seven weeks after her hip replacement.
 
Additional Manufacturer Narrative
The affected cobalt chrome 12/14 taper femoral head, tandem international bipolar and synergy porous stem were not returned for evaluation.Our investigation including a review of complaint history revealed no additional complaints for the listed batches with the same failure mode.The devices were manufactured in 2016 and 2017.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.A clinical evaluation noted that without the explant analysis and based solely on the unidentified copy of the radiological image, the exact cause of the inferolaterally displaced cup/head could not be determined.The non-radiopaque retention ring cannot be identified on the provided image.Although trauma was not reported, it is unknown if it was a precursor to such a significant dislocation.The patient impact was the reported dislocation, revision procedure, expected pain/discomfort, and a post-revision rehabilitation phase.No further medical assessment can be rendered at this time.Without the return of the actual products involved, our investigation of this report is inconclusive.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.Should the device or additional information be received, the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COCR 12/14 FEM HD 22 + 0
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8124323
Report Number1020279-2018-02653
Device Sequence Number2165472
Product Code LPH
UDI-Device Identifier03596010194039
UDI-Public03596010194039
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K963509
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number71302200
Device Lot Number17HM11285
Initial Date Received by Manufacturer 11/04/2018
Supplement Date Received by Manufacturer11/04/2018
Initial Report FDA Received Date12/03/2018
Supplement Report FDA Received Date02/05/2019
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age79 YR
-
-