| Model Number |
1884080EM |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
A health care provider (hcp) reported via manufacturer representative that a blade fell apart, pre-operatively.There was no patient impact.On follow up, it was confirmed that the blade fell apart when they opened the packaging, a new blade was provided to complete the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Visually, the tracker hub became detached from the outer hub which would have resulted in the reported event.The components are to be bonded together with adhesive.There was a residue consistent with adhesive on the mounting surfaces.There was an out of specification condition identified as it relates to the complaint.If information is provided in the future, a supplemental report will be issued.
|
| |
|