• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOLOGIC, INC. NOVASURE IMPEDANCE CONTROLLED EA SYSTEM; UTERINE ABLATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOLOGIC, INC. NOVASURE IMPEDANCE CONTROLLED EA SYSTEM; UTERINE ABLATION DEVICE Back to Search Results
Model Number NS2007
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Uterine Perforation (2121)
Date of Event 11/08/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device has not yet been returned therefore, a failure analysis of the complaint device cannot be completed.Device history record (dhr) review was conducted for the reported identification number.The lot was released meeting all qa specifications.Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that during a novasure endometrial ablation procedure a uterine perforation was noted as "two small holes near the left and right of the fundus." the perforation was first recognized when the cia (cavity integrity assessment) test failed and then by hysteroscope view of the cavity post novasure device removal.No medical intervention was needed and the patient was discharged home the same day.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOVASURE IMPEDANCE CONTROLLED EA SYSTEM
Common Device Name
UTERINE ABLATION DEVICE
Manufacturer (Section D)
HOLOGIC, INC.
250 campus drive
marlborough MA 01752
MDR Report Key8124850
Report Number1222780-2018-00246
Device Sequence Number16206340
Product Code MNB
UDI-Device Identifier35420045501137
UDI-Public35420045501137
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P010013
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Operator of Device Health Professional
Device Expiration Date09/07/2019
Device Model NumberNS2007
Device Catalogue NumberNS2007
Device Lot Number18H07R
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/08/2018
Initial Report FDA Received Date12/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
RADIO FREQUENCY CONTROLLER: SN UNKNOWN
Outcome Attributed to Adverse Event Other;
-
-