| Model Number |
M00558430 |
| Medical Device Problem Codes |
Material Puncture/Hole (1504); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient's exact age is unknown; however it was reported that the patient was over the age of 18.(b)(6).(b)(4).Investigation result: a visual examination of the complaint device revealed that the exit marker bunched up.Functional analysis was performed, and the balloon was inflated; however, the balloon would not hold pressure due to a pinhole found in the body of the balloon.This failure is likely due to an expected or random component failure without any design or manufacturing issue.Therefore, the most probable root cause is cause traced to component failure.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a cre wireguided dilatation balloon was used in the biliary duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the balloon could not be inflated.The procedure was completed with another cre wireguided dilatation balloon.There were no patient complications reported as a result of this event.Investigation results revealed that the balloon had a pinhole and the exit marker bunched up; therefore, this is now an mdr reportable event.
|
| |
|