| Model Number |
MMT-751PNAL |
| Medical Device Problem Codes |
Mechanical Problem (1384); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
10/31/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that they had experienced high blood glucose and insulin pump received motor error.The customer¿s blood glucose level was 1770 mg/dl.The customer reported that they received a motor error alarm during bolus.Customer was able to successfully clear the alarm.Customer was able to complete pump rewind.The customer reported that the drive support cap was normal.The customer was advised the pump will need to be replaced.The customer was advised to discontinue use of the pump and was advised to refer to back up plan, per health care professional instructions.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump passed displacement test, rewind test, basic occlusion test, prime test, excessive no delivery test and occlusion test.No motor error alarm noted.The motor was tested outside of the device and passed.
|
| |
|