| Model Number |
9731965 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
08/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The probe was returned to the manufacturer for analysis.Analysis found that when the markers attached and fully seated, the returned probe displayed a high geometry error but a good divot error.The support arms for markers #3 and #4 were bent causing the high geometry error.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation device being used outside of a procedure.It was reported that the touch n go probe was bent.There was no patient present when this issue was observed.
|
| |
|