| Catalog Number |
48375094 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The cage broke during surgery, broken fragments were retrieved.Replacement cage was used to complete the surgery.10 minute surgical delay was reported.No adverse consequences to the patient has been reported.
|
| |
|
Event or Problem Description
|
|
The cage broke during surgery, broken fragments were retrieved.Replacement cage was used to complete the surgery.10 minute surgical delay was reported.No adverse consequences to the patient has been reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Method: visual inspection, product history review, complaint history review, nc/capa history review, labelling review, risk assessment.Result: the reported event was confirmed via visual inspection of returned product.Manufacturing records were reviewed for the corresponding lot and no relevant issues were identified.It was also reported that the cage was impacted with straight impactor via high amount of force.Conclusion: the root cause of the reported event was determined to be user error- too much impaction force.
|
| |
|
Search Alerts/Recalls
|