| Catalog Number |
213524 |
| Medical Device Problem Codes |
Break (1069); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
Pin (213524) tip broke off in tibia upon insertion.Surgeon attempted to put pin in tibial crest of male patient with dense bone (according to surgeon).Case type: pka.Post x-ray attached.
|
| |
|
Event or Problem Description
|
|
Pin (213524) tip broke off in tibia upon insertion.Surgeon attempted to put pin in tibial crest of male patient with dense bone (according to surgeon).Case type: pka.Post x-ray.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: pin (213524) tip broke off in tibia upon insertion.Surgeon attempted to put pin in tibial crest of male patient with dense bone (according to surgeon).Case type: pka , post x-ray.Product evaluation and results: the product was unavailable for inspection as the product was not returned.The post op x-ray provided shows a radiopaque object in the typical location and similar size for the bone pin tip.Product history review: review of the product history records indicate (b)(4) devices were manufactured under lot no 48011117 and accepted on 12-21-17.Review of the product history records indicate (b)(4) devices were manufactured under lot no 48011117 and accepted on 12-22-17.Review of the product history records indicate (b)(4) devices were manufactured under lot no 48011117 and accepted on 12-22-17.Complaint history review: a review of complaints in trackwise related to p/n 213524, l/n 48011117 shows no additional complaints under investigation related to the failure in this investigation.Conclusions: the failure mode could not be confirmed because the part was not available for evaluation.If device and/or additional information become available, this investigation will be reopened.Corrective action/preventive action: a review of stryker¿s nc/capa database indicated there have been an ncs 1470754 and capa 1480798 are associated with the product and failure mode reported in this event.
|
| |
|