| Model Number |
DBP-150SOLID145 |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
Vascular Dissection (3160)
|
| Date of Event |
11/09/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Analysis of the reported device is in progress.A supplemental report will be submitted when the device analysis is completed.(b)(4).
|
| |
|
Event or Problem Description
|
|
During a peripheral atherectomy procedure using a csi orbital atherectomy device (oad), a dissection occurred.During treatment of a lesion in the superficial femoral artery (sfa), the device stopped spinning.Upon removal of the oad, imaging was performed and a type a dissection was noted.The guide wire was observed to be in a false lumen, and the location was corrected.Balloon angioplasty was performed and an unplanned stent was placed to resolve the dissection.The patient was in stable condition following the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported oad was received for analysis.Adhered biological material was observed on the driveshaft filars and crown.The material did not prevent a guide wire from passing through the device, and examination of the areas of adhered tissue did not reveal any damage which may have contributed to the accumulation.When tested, the device functioned as intended with no anomalies observed.It is possible that the device became wrapped in biological material which led to a stall condition, however this could not be conclusively confirmed through analysis.The device history record for this oad lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.(b)(4).
|
| |
|
Search Alerts/Recalls
|