• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. PROXIMATE*RL TITAN HEAVY WIRE; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. PROXIMATE*RL TITAN HEAVY WIRE; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number TLH90
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch # unk.The device history records were reviewed and the manufacturing criteria were met prior to the release of this lot.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported that during a rectum procedure, staples malformed.During the open rectum, after fired, noted the staples malformed.Another device was used to complete the surgery.There were no adverse consequences to the patient.
 
Additional Manufacturer Narrative
(b)(4).Batch # p58v1y.The analysis results showed that the tlh90 device was received with the hook shroud opened by the seam and with a cartridge loaded on the device.The cartridge was returned fully fired.No functional test could be performed due to the condition of the device.The device was disassembled to verify the condition of the internal components and the device was noted to have the lockout slide tip and the adjusting knob damaged.The damage to the lockout slid tip and knob is consistent with an excessive force applied to the rotating knob when trying to dial down a locked device, causing the damage to the knob.It should be noted that the instrument has been designed with a lockout feature, which during the first application, prevents turning the adjusting knob unless the retaining pin is pushed completely forward.Please reference the instructions for use for additional information.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key8126292
Report Number3005075853-2018-14803
Device Sequence Number1579130
Product Code GDW
UDI-Device Identifier10705036013242
UDI-Public10705036013242
Combination Product (Y/N)N
PMA/510(K) Number
K020779
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/27/2022
Device Catalogue NumberTLH90
Device Lot NumberP94F7Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2018
Initial Date Received by Manufacturer 11/11/2018
Supplement Date Received by Manufacturer12/06/2018
Initial Report FDA Received Date12/03/2018
Supplement Report FDA Received Date12/21/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-